Go to main content

Background: This study aimed to evaluate the within- and between-session reliability and the convergent validity of a mobile application for measuring discrete and continuous parameters of spinal kinematics in real-life settings. As a secondary objective, we explored the impact of the number of repetitions on reliability. Methods: Spinal kinematics was measured using the mobile application Digiback during two different sessions in 30 asymptomatic participants (50% male; 36.7 ± 14.07 years old). Within- and between-session reliability of discrete parameters was calculated using the Intraclass Correlation Coefficient (ICC), while reliability of the continuous data was calculated using pointwise ICC values over the time-normalized amplitude data curves. Convergent validity was analysed against one inertial sensor attached to the upper lumbar spine. Results: Reliability of discrete parameters was demonstrated by ICCs above 0.75 for most parameters. Reducing the number of repetitions to three had a minimal impact on ICCs. Integrated ICCs of continuous parameters varied between 0.34 and 0.79 for the whole curve, and from 0.64 to 0.88 for the mid portion (20-80%) of the movement curve. Convergent validity of discrete parameters varied across movements, with higher validity for sagittal-plane angular amplitude and velocity. Conclusions: This study supports the reliability of measuring discrete and continuous sagittal-plane parameters of spinal kinematics with a smartphone in asymptomatic individuals. Reducing the number of repetitions from five to three had minimal impact on reliability, supporting clinical feasibility. Convergent validity was demonstrated only for discrete parameters in the sagittal plane.